Abstract
Background: This work evaluates the accuracy and agreement between the FreeStyle Libre sensor (FSL) and an off-label converted real-time continuous glucose monitor (c-rtCGM) device consisting of the MiaoMiao transmitter and the xDrip+ application which can be coupled to the FSL. Methods: Four weeks of glucose data were collected from 21 participants with type 1 diabetes using the c-rtCGM and FSL: two weeks with a single initial calibration (uncalibrated) and two weeks with a daily calibration (calibrated). Accuracy and agreement evaluation included mean absolute relative difference (MARD), the %20/20 rule, Bland-Altman plots, and the Consensus Error Grid analysis. Results: Values reported by the c-rtCGM system compared with the FSL resulted in an overall MARD of 12.06% and 84.71% of the results falling within Consensus Error Grid Zone A when the device is calibrated. For uncalibrated devices, an overall MARD of 17.49% was obtained. Decreased accuracy was shown in the hypoglycemic range and for rates of change greater than 2 mg/dL/min. The between-device bias also incremented with increasing glucose values. Conclusion: Measurements recorded by the c-rtCGM were found to be accurate when compared with FSL data only when performing daily c-rtCGM device calibrations. High drops in accuracy and agreement between devices occurred when the c-rtCGM was not calibrated.
| Original language | English |
|---|---|
| Journal | Journal of diabetes science and technology |
| DOIs | |
| State | Accepted/In press - 2022 |
Bibliographical note
Publisher Copyright:© 2022 Diabetes Technology Society.
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- FreeStyle Libre
- MiaoMiao transmitter
- continuous glucose monitoring
- intermittent scanned glucose monitoring
- xDrip+ software
Types Minciencias
- Artículos de investigación con calidad A1 / Q1
Fingerprint
Dive into the research topics of 'Accuracy of an Off-Label Transmitter and Data Manager Paired With an Intermittent Scanned Continuous Glucose Monitor in Adults With Type 1 Diabetes'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver