T1 - Accuracy of an Off-Label Transmitter and Data Manager Paired With an Intermittent Scanned Continuous Glucose Monitor in Adults With Type 1 Diabetes
AU - Villa-Tamayo, María F.
AU - Builes-Montaño, Carlos E.
AU - Ramirez-Rincón, Alex
AU - Carvajal, Javier
AU - Rivadeneira, Pablo S.
N1 - Publisher Copyright:
© 2022 Diabetes Technology Society.
PY - 2022
Y1 - 2022
N2 - Background: This work evaluates the accuracy and agreement between the FreeStyle Libre sensor (FSL) and an off-label converted real-time continuous glucose monitor (c-rtCGM) device consisting of the MiaoMiao transmitter and the xDrip+ application which can be coupled to the FSL. Methods: Four weeks of glucose data were collected from 21 participants with type 1 diabetes using the c-rtCGM and FSL: two weeks with a single initial calibration (uncalibrated) and two weeks with a daily calibration (calibrated). Accuracy and agreement evaluation included mean absolute relative difference (MARD), the %20/20 rule, Bland-Altman plots, and the Consensus Error Grid analysis. Results: Values reported by the c-rtCGM system compared with the FSL resulted in an overall MARD of 12.06% and 84.71% of the results falling within Consensus Error Grid Zone A when the device is calibrated. For uncalibrated devices, an overall MARD of 17.49% was obtained. Decreased accuracy was shown in the hypoglycemic range and for rates of change greater than 2 mg/dL/min. The between-device bias also incremented with increasing glucose values. Conclusion: Measurements recorded by the c-rtCGM were found to be accurate when compared with FSL data only when performing daily c-rtCGM device calibrations. High drops in accuracy and agreement between devices occurred when the c-rtCGM was not calibrated.
AB - Background: This work evaluates the accuracy and agreement between the FreeStyle Libre sensor (FSL) and an off-label converted real-time continuous glucose monitor (c-rtCGM) device consisting of the MiaoMiao transmitter and the xDrip+ application which can be coupled to the FSL. Methods: Four weeks of glucose data were collected from 21 participants with type 1 diabetes using the c-rtCGM and FSL: two weeks with a single initial calibration (uncalibrated) and two weeks with a daily calibration (calibrated). Accuracy and agreement evaluation included mean absolute relative difference (MARD), the %20/20 rule, Bland-Altman plots, and the Consensus Error Grid analysis. Results: Values reported by the c-rtCGM system compared with the FSL resulted in an overall MARD of 12.06% and 84.71% of the results falling within Consensus Error Grid Zone A when the device is calibrated. For uncalibrated devices, an overall MARD of 17.49% was obtained. Decreased accuracy was shown in the hypoglycemic range and for rates of change greater than 2 mg/dL/min. The between-device bias also incremented with increasing glucose values. Conclusion: Measurements recorded by the c-rtCGM were found to be accurate when compared with FSL data only when performing daily c-rtCGM device calibrations. High drops in accuracy and agreement between devices occurred when the c-rtCGM was not calibrated.
KW - FreeStyle Libre
KW - MiaoMiao transmitter
KW - continuous glucose monitoring
KW - intermittent scanned glucose monitoring
KW - xDrip+ software
UR - http://www.scopus.com/inward/record.url?scp=85141694450&partnerID=8YFLogxK
U2 - 10.1177/19322968221133405
DO - 10.1177/19322968221133405
M3 - Artículo en revista científica indexada
C2 - 36281579
AN - SCOPUS:85141694450
SN - 1932-2968
JO - Journal of diabetes science and technology
JF - Journal of diabetes science and technology
ER -